Electro surgical unit

 PURPOSE •ESU are used for surgical cutting and for controlling bleeding by causing coagulation “hemostasis” at the surgical site. •It is “high frequency diathermy” involves the transference of electrical energy into heat which ,when impact to the bodily tissues, will heat the normal cell fluid, eventually through it is boiling point. •What dose “DIATHERMY” mean? DIA = through THERMO = temperature heat

 Principle • ESU machine is an alternating current source that operates at radio frequency “RF”. • A high frequency current flowing through active electrode “high current concentration”. • Cell ruptured-fumes or evaporates. • Return path through dispersive electrode “low current and heat dissipates”. • Patient is included in circuit. • Current concentration or density depends on the size of the area through which the current flows. • HF generation can be activated by a foot switch or finger switch on the surgical handle.

System Components

Electrosurgery Unit

Dispersive Electrode

Active Electrode

Active electrode

RF Generator

Body

Patient electrode

..\ESU Principle.ppt

 Effect of RF on cell •

When a high frequency current applied to the tissues the tissues gets turn apart and get the following effect: 1. Thermal effect. 2. Electrolytic effect. 3. Faradic effect.

 Operating frequency and typical value • Operating frequency of solid state surgical diathermy machine is 300 KHz –to-3MHz • Monopolar :CUT “0-to-350”watts for load 5 COAG”0-to-100” watts • Bipolar : CUT “0-to-50” watts COAG “0-to-10” watts

 Types of ESU Spark gap generator – 1924 • Still in use today “urology , open- heart surgery”. • Spark gap / vacuum-tube device use spark –gap circuits to generate high frequency waveforms. • It is not offer the same safety feature as solid state unites.

 Recent technology • Argon/argon enhanced technology. • It uses a computer-controlled tissue feedback system that senses tissue impedance and automatically adjusts the current and output voltage to maintain a constant surgical effect . • It is reduces the need to adjust power setting for different types of tissue. • It is also gives improved performance at low power setting to reduce the risk of patient injury.

 Solid state device - 1968 • It is more recent and more prevalently used technology. • Contain oscillator circuits and transistor-based amplifier that vary the frequency and modify the shape of the line signal to create an array of different waveforms for pure CUT,COAG, BLEEND . • Highly safety.

OK!, but how can I control so accurately the temperatures in the tissues ?

 Modes of Electro surgery Monopolar surgery

Bipolar surgery

 Modes • • • •

of electrosurgery

Electrotomy/cutting desiccation / coagulation Blend fulgurations

General Block Diagram Audio tone generator Front panel

Main CPU

LED display

Logic/ Control Board Rely board

Power supply

Isolation transformer

Power O/P or RF board

Foot switch

Patient return electrode

RF output

View more...