Service Manual

9650-0308-01 Rev. A March 2005

An issue date and revision level for this manual appear on the front cover. If more than three years have elapsed since this date, contact ZOLL Medical Corporation to determine if additional product information updates are available. ZOLL and stat•padz are registered trademarks, and AED Plus, AED Pro, CPR-D•padz, pedi•padz, pro•padz, PowerCharger, Preconnect, RescueNet, Smart Alarms, and v•pak are trademarks of ZOLL Medical Corporation. All other trademarks and registered trademarks are property of their respective owners. Copyright © 2005 ZOLL Medical Corporation. All rights reserved.

Table of Contents Preface Related Documentation .........................................................................................................iv Conventions ...........................................................................................................................iv Symbols ..................................................................................................................................v Safety Considerations............................................................................................................vi Warranty ................................................................................................................................vi Technical Service.................................................................................................................. vii Returning a Unit for Service .......................................................................................... vii

Chapter 1

Maintenance Tests

Preparations for Testing ...................................................................................................... 1-2 Test Equipment ................................................................................................................... 1-2 1.0 Physical Inspection.................................................................................................. 1-3 2.0 Preventive Maintenance Test .................................................................................. 1-4 3.0 Manual Self-Test ..................................................................................................... 1-9

Chapter 2

Troubleshooting

Troubleshooting the AED Plus Unit .................................................................................... 2-2 AED Plus Error Log Messages ........................................................................................... 2-4

Appendix A

Maintenance Tests Checklist

Instructions .........................................................................................................................A-2 ZOLL AED Plus Maintenance Tests Checklist ....................................................................A-3

Index

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

i

TABLE OF CONTENTS

ii

www.zoll.com

9650-0308-01 Rev. A

Preface ZOLL Medical Corporation provides this manual for technicians who service the AED Plus™ device or diagnose malfunctions. This manual contains the following sections: Preface — Provides safety warnings and an overview of the manual’s contents. Review this section thoroughly before servicing an AED Plus unit. Chapter 1, "Maintenance Tests" — Provides the recommended procedures for evaluating the condition and performance of the AED Plus unit. Chapter 2, "Troubleshooting" — Describes possible problems and solutions and lists the error log messages that indicate the unit requires technical service. Appendix A, "Maintenance Tests Checklist" — Provides a checklist for recording the results of maintenance tests. Note:

9650-0308-01 Rev. A

The AED Plus contains no user-replaceable parts.

ZOLL AED Plus Service Manual

iii

PREFACE

Related Documentation In addition to this service manual, the following manuals are available for the AED Plus unit: • AED Plus Operator’s Guide (9650-0300-01) — Describes what the rescuer sees, hears, and

does while operating the unit. • AED Plus Administrator’s Guide (9650-0301-01) — Describes the features and functions of the unit and the procedures for configuring and maintaining it. • AED Plus Simulator Operator’s Guide (9651-0800-01) — Describes the AED Plus simulator, which you can use to test the AED Plus unit’s response to various simulated cardiac rhythms.

Conventions This manual uses uppercase italics for audible prompts and for text messages displayed on the screen (for example DON’T TOUCH PATIENT, ANALYZING).

!

Within test or troubleshooting procedures, this symbol indicates an audible prompt or message; for example:

!

UNIT OK

WARNING!

Warning statements alert you to conditions or actions that can result in personal injury or death.

Caution

Caution statements alert you to conditions or actions that can result in damage to the unit.

iv

www.zoll.com

9650-0308-01 Rev. A

Symbols

Symbols Symbols used in this manual or on the equipment include the following: Class II equipment Defibrillation-protected Type BF patient connection ATTENTION: Refer to the operator’s guide for more information

WARNING: Dangerous voltage

New battery cells

Not new battery cells

Push button

Do not push button

Recommended manufacturer VARTA

Do not use this manufacturer

Unit equipped to treat adult and pediatric patients

Conformité Européenne Complies with medical device directive 93/42/EEC.

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

v

PREFACE

Safety Considerations Only qualified personnel should service or disassemble an AED Plus unit. Before using this device, read the AED Plus Administrator’s Guide. Before servicing or disassembling any equipment, review these safety considerations and read this manual carefully.

WARNING!

The AED Plus unit can generate as much as 2250 volts with sufficient current to cause lethal shocks. Before discharging the defibrillator, warn everyone near the equipment to “STAND CLEAR.” Use only Type 123A lithium manganese dioxide batteries from manufacturers recommended by ZOLL. Discard used batteries properly. Do not recharge, disassemble, or dispose of batteries in fire. Batteries might explode if mistreated. This device is protected against interference from radio frequency emissions typical of the two-way radios and cellular telephones used in emergency service and public safety activities. You should assess the device’s performance in your typical operating environment to determine the likelihood of radio frequency interference (RFI) from high-power sources.

Warranty For warranty information, refer to the AED Plus Administrator’s Guide. To maintain the warranty, strictly follow the instructions and procedures in this service manual. Unit failure that is attributable to the use of accessories that are not manufactured by ZOLL might void the ZOLL warranty.

vi

www.zoll.com

9650-0308-01 Rev. A

Technical Service

Technical Service If the AED Plus unit requires service, contact the ZOLL Technical Service Department or your local distributor. Telephone:

1-800-348-9011 (U.S.A. customers only) 1-978-421-9655

Fax:

1-978-421-0010

When requesting service, please provide the following information to the service representative: • • • • •

Unit serial number Description of the problem Department where the equipment is used and the name of the person to contact Purchase order to allow tracking of loan equipment Purchase order for a unit with an expired warranty

Returning a Unit for Service Before sending a unit to the ZOLL Technical Service Department for repair, obtain a service request (SR) number from the service representative. Remove all batteries from the unit. Pack the unit in its original container or equivalent packaging. Be sure the assigned service request number appears on the package. For customers

Return the unit to

In the U.S.A.

ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 Attention: Technical Service Dept. [ SR number ]

In Canada

ZOLL Medical Canada Unit #15 5266 General Road Mississauga, Ontario L4W 1Z7 Attention: Technical Service Dept. [ SR number ] Telephone: 1-866-442-1011

In other locations

The nearest authorized ZOLL Medical Corporation representative. To locate an authorized service center, contact the International Sales Department at ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 Telephone: 1-978-421-9655

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

vii

PREFACE

viii

www.zoll.com

9650-0308-01 Rev. A

Chapter 1 Maintenance Tests This chapter describes the procedures for routine testing of the AED Plus unit. A test typically consists of multiple steps, each of which requires a pass/fail rating. For each step, perform the specified action and then verify that the expected results occur. If all expected results occur, assign a “Pass” rating to the step; otherwise, assign it a “Fail” rating. This chapter includes the following inspection and test procedures: • “1.0 • “2.0 • “3.0

9650-0308-01 Rev. A

Physical Inspection” on page 1-3 Preventive Maintenance Test” on page 1-4 Manual Self-Test” on page 1-9

ZOLL AED Plus Service Manual

1–1

CHAPTER 1

MAINTENANCE TESTS

Preparations for Testing • Gather the required equipment listed in the next section. • Perform all the steps within each test procedure in the order specified.

If testing results indicate that the batteries must be replaced, you will need a set of 10 fully charged Type 123A batteries.

Test Equipment For some tests, you need one or more of these items: • ZOLL universal adapter cable • Fluke® Impulse 4000 Defibrillator/Pacer Performance Analyzer (Fluke Biomedical/

DNI Nevada) with 1.06 software or higher The following items are optional: • ZOLL AED Plus Simulator —

When connected to an AED Plus unit, the simulator can send it various simulated cardiac rhythms (ventricular fibrillation, normal sinus rhythm, or asystole) and optional simulated chest compression readings (rate and depth), so that you can test the unit’s response. • ZOLL Administration Software (ZAS) —

When installed on a personal computer or personal digital assistant that has an IrDA® (wireless infrared) interface, the ZAS application can communicate with an AED Plus unit in nonrescue mode to retrieve its usage history, hardware and software configuration, and error log. You can also use ZAS to view or modify the unit’s configurable settings and to retrieve event data. For ZOLL part numbers, refer to the accessories list in the AED Plus Administrator’s Guide.

1–2

www.zoll.com

9650-0308-01 Rev. A

Physical Inspection

1.0

Physical Inspection Use this procedure to ensure that the unit shows no signs of damage or excessive wear.

Tools Needed None.

Test Setup None.

Test Procedure Observe the following:

Pass

Fail

1.1

Is the unit clean?

!

!

1.2

Are all signs of wear reasonable and not excessive?

!

!

1.3

Are the top panel and housing undamaged and free of cracks?

!

!

1.4

Are input connectors clean and undamaged?

!

!

1.5

If electrodes are preconnected, is the cable free from damage, are the pads sealed within the packaging, and are the electrodes within the expiration date shown on the packaging?

!

!

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

1–3

CHAPTER 1

2.0

MAINTENANCE TESTS

Preventive Maintenance Test Use this test to verify that the AED Plus unit is working properly and ready for use. For an alternative test using the AED Plus Simulator, refer to the AED Plus Administrator’s Guide. Note:

To verify that the delivered energy for the first, second, and third shocks corresponds to the unit’s preconfigured adult energy levels, use the ZOLL Administration Software (ZAS) to obtain the current configuration of the unit.

Tools Needed • Fluke Impulse 4000 Defibrillator/Pacer Performance Analyzer or equivalent • Universal adapter cable Test Setup • Turn off the AED Plus unit, remove the cover, and unplug the preconnected electrodes. Keep the electrodes so they can be reconnected later. • Connect the universal adapter cable to the AED Plus unit and to the analyzer. • Turn on the analyzer and set its function to VF (ventricular fibrillation). Variations in Messages When verifying messages in the test results, please be aware of the following issues: • The wording of some messages is configurable. The following table shows equivalent

messages. CHECK RESPONSIVENESS

CHECK PATIENT

CHECK CIRCULATION

CHECK PULSE

IF NO CIRCULATION START CPR

IF NO PULSE START CPR

IF NO CIRCULATION CONTINUE CPR

IF NO PULSE CONTINUE CPR

TREATMENT ADVISED

SHOCK ADVISED

NO TREATMENT ADVISED

NO SHOCK ADVISED

PRESS TREATMENT BUTTON

PRESS SHOCK BUTTON

RELEASE TREATMENT BUTTON

RELEASE SHOCK BUTTON

TREATMENT DELIVERED

SHOCK DELIVERED

NO TREATMENT DELIVERED

NO SHOCK DELIVERED

• The message that reports the detected electrode type can be either ADULT PADS or

PEDIATRIC PADS. • Some messages are issued only if the unit is configured to do so. If an optional message is not issued during a test, do not consider this a failure. • In some cases a text message on the display screen is abbreviated or otherwise differs slightly from the associated voice message.

1–4

www.zoll.com

9650-0308-01 Rev. A

Preventive Maintenance Test Test Procedure Test

Action

Expected Result(s)

2.1

Turn on the AED Plus unit.

!

UNIT OK

!

ADULT PADS

!

DON’T TOUCH PATIENT, ANALYZING

!

TREATMENT ADVISED (or SHOCK ADVISED)

!

DON’T TOUCH PATIENT.

!

PRESS TREATMENT BUTTON (or PRESS SHOCK BUTTON)

2.2

Press the Treatment button (first shock).

• Analyzer indicates a defibrillator discharge at the configured first-shock energy level (120, 150 or 200 joules ±15%).

!

Pass

Fail

!

!

!

!

TREATMENT DELIVERED (or SHOCK DELIVERED)

• An updated shock count appears on the display (SHOCKS: 1).

9650-0308-01 Rev. A

!

DON’T TOUCH PATIENT, ANALYZING

!

TREATMENT ADVISED (or SHOCK ADVISED)

!

DON’T TOUCH PATIENT.

!

PRESS TREATMENT BUTTON (or PRESS SHOCK BUTTON)

ZOLL AED Plus Service Manual

1–5

CHAPTER 1

MAINTENANCE TESTS Test

Action

Expected Result(s)

2.3

Press the Treatment button (second shock).

• Analyzer indicates a defibrillator discharge at the configured second-shock energy level (120, 150 or 200 joules ±15%).

!

Pass

Fail

!

!

!

!

TREATMENT DELIVERED (or SHOCK DELIVERED)

• An updated shock count appears on the display (SHOCKS: 2).

2.4

Press the Treatment button (third shock).

!

DON’T TOUCH PATIENT, ANALYZING

!

TREATMENT ADVISED (or SHOCK ADVISED)

!

DON’T TOUCH PATIENT.

!

PRESS TREATMENT BUTTON (or PRESS SHOCK BUTTON)

• Analyzer indicates a defibrillator discharge at the configured third-shock energy level (120, 150 or 200 joules ±15%).

!

TREATMENT DELIVERED (or SHOCK DELIVERED)

• An updated shock count appears on the display (SHOCKS: 3). These three prompts are optional: — OPEN AIRWAY — CHECK BREATHING — GIVE TWO BREATHS

1–6

!

CHECK CIRCULATION (or CHECK PULSE)

!

IF NO CIRCULATION START CPR (or IF NO PULSE START CPR)

www.zoll.com

9650-0308-01 Rev. A

Preventive Maintenance Test Test

Action

Expected Result(s)

Pass

Fail

!

!

!

!

!

!

Turn off the AED Plus unit. Set the analyzer to NSR (normal sinus rhythm). 2.5

Turn on the AED Plus unit.

!

UNIT OK

!

ADULT PADS

!

DON’T TOUCH PATIENT, ANALYZING.

!

NO TREATMENT ADVISED (or NO SHOCK ADVISED)

These three prompts are optional: — OPEN AIRWAY — CHECK BREATHING — GIVE TWO BREATHS

2.6

Lift the puck that is attached to the universal adapter cable. Move the puck up and down very slightly (no more than 1 inch or 2 cm) and continue to do so.

!

CHECK CIRCULATION (or CHECK PULSE)

!

IF NO CIRCULATION START CPR (or IF NO PULSE START CPR)

• Unit emits spaced tones as a guide for performing chest compressions at the proper rate.

!

PUSH HARDER

If the unit has a display, prevent the bar on the chest compression gauge from extending into the area between the lower two lines. 2.7

Increase the vertical motion of the puck (about 2 to 3 inches or 5 to 8 cm).

• Unit emits spaced tones as a guide for performing chest compressions at the proper rate.

If the unit has a display, make sure the bar on the chest compression gauge extends into the area between the lower two lines.

!

The following prompt is optional: — IF NO CIRCULATION CONTINUE CPR (or IF NO PULSE CONTINUE CPR)

!

9650-0308-01 Rev. A

GOOD COMPRESSIONS

STOP CPR

ZOLL AED Plus Service Manual

1–7

CHAPTER 1

MAINTENANCE TESTS Test

Action

Expected Result(s)

Pass

Fail

!

!

!

!

!

!

Put the puck down and turn off the AED Plus unit. Disconnect the universal adapter cable from the AED Plus unit. 2.8

Wait 10 seconds and then turn on the AED Plus unit.

!

PLUG IN CABLE

This prompt is optional: — STAY CALM

!

CHECK RESPONSIVENESS (or CHECK PATIENT)

This prompt is optional: — CALL FOR HELP These three prompts are optional: — OPEN AIRWAY — CHECK BREATHING — GIVE TWO BREATHS

2.9

Check the expiration date on the previously connected defibrillation electrodes.

!

CHECK CIRCULATION (or CHECK PULSE)

!

PLUG IN CABLE

!

ADULT PADS (or PEDIATRIC PADS)

If the electrodes have not expired, reconnect them to the unit; otherwise, connect a fresh package of defibrillation electrodes. Turn off the AED Plus unit. 2.10

1–8

After 90 seconds, check the Status indicator on the unit.

• The Status indicator shows a green check.

www.zoll.com

9650-0308-01 Rev. A

Manual Self-Test

3.0

Manual Self-Test Use this test to initiate the unit’s self-diagnostic procedure, which verifies the following: • Battery usage since last reset. • Defibrillation electrodes are properly connected. • ECG signal acquisition and processing circuitry is functional. • Defibrillator charge and discharge circuitry (operational at 2 joules) is functional. • Microprocessor hardware and software are functional. • CPR monitoring circuitry and compression depth sensor are functional (when CPR-D•padz

are attached). • Audio output circuitry is functional. • Display screen (if present) is functional.

Tools Needed • None. Test Setup • Turn off the AED Plus unit. • If they are not already in place, preconnect defibrillation electrodes to the unit. Test Procedure Test

Action

Expected Result(s)

3.1

Press and hold the ON/OFF button for at least 5 seconds.

• Within 10 seconds after the unit is turned on, the Status indicator displays a green check.

Pass

Fail

!

!

• On the top panel, the indicator lights for the rescue actions illuminate in sequence.

!

UNIT OK

If an IrDA connection is established, turn off the unit; otherwise, the unit shuts down automatically.

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

1–9

CHAPTER 1

1–10

MAINTENANCE TESTS

www.zoll.com

9650-0308-01 Rev. A

Chapter 2 Troubleshooting This chapter describes common technical issues encountered during routine maintenance or after a malfunction of the AED Plus unit. This chapter also lists critical AED Plus error log messages that indicate the unit requires technical service. This chapter contains the following sections: • “Troubleshooting the AED Plus Unit” on page 2-2 • “AED Plus Error Log Messages” on page 2-4

If this chapter does not help you resolve the problem, call the ZOLL Technical Service Department for assistance. For contact information, refer to page vii.

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

2–1

CHAPTER 2

TROUBLESHOOTING

Troubleshooting the AED Plus Unit The following tables include common issues and their solutions. First, try the recommendations given under “Operator Action.” If these steps do not remedy the problem, follow the suggestions under “Technical Action.” For information on contacting ZOLL Technical Service, refer to page vii. Table 2-1.

AED Plus Troubleshooting

Problem

Operator Action

Technical Action

Status indicator shows a red “X”.

— Press and hold the ON/OFF button for at least 5 seconds to perform a self-test.

If the unit continues to fail, remove it from service and contact Technical Service.

— Verify that the electrode cable is properly plugged into the patient cable connector. — Replace the electrodes.

!

UNIT FAILED

—Turn the unit off and then press and hold the ON/OFF button for at least 5 seconds to perform a self-test.

If the unit continues to fail, remove it from service and contact Technical Service.

— Verify that the electrode cable is properly plugged into the patient cable connector. — Replace the electrodes. Unit beeps while turned off.

— Verify that the electrode cable is properly plugged into the patient cable connector.

If the unit continues to beep, remove it from service and contact Technical Service.

—Turn the unit off and then press and hold the ON/OFF button for at least 5 seconds to perform a self-test. Verify that a green check appears in the Status indicator.

!

CHANGE BATTERIES

!

PLUG IN CABLE

!

ANALYSIS HALTED KEEP PATIENT STILL

!

NO SHOCK DELIVERED (or NO TREATMENT DELIVERED)

— Replace all batteries with new ones. When prompted, press the battery reset button.

If the message persists, remove the unit from service and contact Technical Service.

— Ensure that the electrode cable is properly plugged into the patient cable connector.

If the message persists, remove the unit from service and contact Technical Service.

— Keep the patient still during ECG analysis. If transporting the patient by stretcher or vehicle, stop all patient movement during analysis.

If the message persists, remove the from service and contact Technical Service.

The Treatment button was not pressed or an internal error occurred.

If an internal error occurred (no shock was delivered when theTreatment button was pressed properly), remove the unit from service and contact Technical Service.

— When prompted to press the button, do so within 30 seconds.

2–2

www.zoll.com

9650-0308-01 Rev. A

Troubleshooting the AED Plus Unit Table 2-1.

AED Plus Troubleshooting (continued)

Problem

Operator Action

Technical Action

!

— Release the Treatment button. Do not press the button until the charge-ready tone sounds and the button begins flashing.

If the message persists, remove the unit from service and contact Technical Service.

RELEASE SHOCK BUTTON (or RELEASE TREATMENT BUTTON)

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

2–3

CHAPTER 2

TROUBLESHOOTING

AED Plus Error Log Messages The AED Plus unit records internal errors in an error log, which you can retrieve by using a Windows-based personal computer with ZOLL Administration Software (ZAS). To retrieve the error log, follow these steps: Step

Action

1

Turn off the AED Plus unit.

2

Start the ZOLL Administration Software (ZAS) on the computer.

3

Position the AED Plus unit so that its IrDA port has a clear line-of-sight with the computer’s IrDA port and the ports are about 10 to 18 inches (25 to 45 cm) apart.

4

Press and hold the unit’s ON/OFF button for at least 5 seconds to start the unit in nonrescue mode.

5

After the unit issues the message COMMUNICATIONS ESTABLISHED, refer to the ZAS help for instructions on saving the unit’s history file and viewing the included error log.

Note:

After saving the error log file on the computer, you can use ZAS to clear the error log on the AED Plus unit.

Table 2-2 lists the critical error log messages. Important: If any of these messages appears in the error log, contact ZOLL Technical Service as described on page vii. Table 2-2. Error No.

2–4

AED Plus Error Log Messages

Error Message

1

The patient relay is not working properly

2

Attempting to charge the capacitor while the SAFELINE in the CPLD is set - should not charge

3

The capacitor takes too long to charge

4

The H-bridge test failed while the bridge was disabled. There should have been 0 current measured.

5

VCAP1 and VCAP2 are not within a defined range of each other

6

The H-bridge test failed while the bridge was enabled - an invalid current reading was taken during the positive test on the H-bridge

7

The H-bridge test failed while the bridge was enabled - an invalid current reading was taken during the negative test on the H-bridge

8

Defib did not charge during the power-on charge test

9

The Defib task is not receiving A2D vCap1 event samples - needed for charging

10

FnSafety has determined that the charge has dropped below a specified value during the holding of charge.

11

The capacitor failed to charge

12

The capacitor over-charged

13

The capacitor contained an unsafe voltage in an idle state

www.zoll.com

9650-0308-01 Rev. A

AED Plus Error Log Messages Table 2-2. Error No.

AED Plus Error Log Messages (continued)

Error Message

14

Patient impedance experienced a short

15

The A2D is getting non-accurate readings during its test

17

The message queue overflowed

18

PowerOn task has timed out while waiting for a self test result to come back.

19

Calibration faults that occur during a clinical event.

20

An unexpected error has occurred when calling a Nucleus operation.

21

An unexpected error has occurred when calling a Rhapsody operation.

22

An unexpected coding error has occurred.

23

No Language installed - corrupt language.

24

The shock button is pressed/stuck.

25

The ECG Calibration data is invalid

26

The Patient Impedance Calibration data is invalid

27

A critical flash error has occurred

28

The WatchDog is not enabled during clinical event

29

The clock test failed

Note: Error numbers 256 and greater indicate noncritical conditions.

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

2–5

CHAPTER 2

2–6

TROUBLESHOOTING

www.zoll.com

9650-0308-01 Rev. A

Appendix A Maintenance Tests Checklist This appendix provides a sample maintenance checklist for the AED Plus unit. To ensure that the equipment is functioning properly and ready for emergency use, complete the checklist periodically according to local protocols. Your organization can create custom checklists for your specific requirements.

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

A–1

APPENDIX A

MAINTENANCE TESTS CHECKLIST

Instructions To maintain records for maintenance testing, photocopy the checklist shown on the following pages. The checklist includes an entry for each test described in this manual. Use these copies to record the results of your testing and then retain the copies for your records.

A–2

www.zoll.com

9650-0308-01 Rev. A

ZOLL AED Plus Maintenance Tests Checklist

Serial No. Location Date

Use this checklist to record the results of AED Plus maintenance tests and keep it for your records. For a description of each numbered test, refer to Chapter 1, "Maintenance Tests" in the AED Plus Service Manual.

Result of Check: " No action required " Minor problems corrected " Disposable supplies replaced " Major problems identified (unit out of service)

1.0 Physical Inspection

2.0 Preventive Maintenance Test

PASS FAIL 1.1 1.2 1.3 1.4 1.5

Additional Remarks

" " " " "

" " " " "

PASS FAIL 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 2.10

9650-0308-01 Rev. A

" " " " " " " " " "

Tester Signature

3.0 Manual Self-Test PASS FAIL 3.1

" " " " " " " " " "

ZOLL AED Plus Service Manual

" "

A–3

A–4

www.zoll.com

9650-0308-01 Rev. A

Index A

I

Administration software, ZOLL 1-2, 2-4 Analyzer 1-2, 1-4

Impulse 4000 analyzer 1-2, 1-4 Inspection, physical 1-3 IrDA port 1-2

C L

Checklists for maintenance tests A-1 Configurable prompts 1-4 Conventions used in manual iv

Log, error 2-4 M

D

Defibrillator analyzer 1-2, 1-4 Documentation iv E

Error log 2-4

Maintenance tests 1-1 checklists A-1 equipment needed 1-2 prerequisites 1-2 preventive 1-4 Manual self-test 1-9 Manuals, related iv Messages error log 2-4 wording 1-4

H

History file, device 2-4

P

Physical inspection 1-3 Port, IrDA 1-2 Preventive maintenance test 1-4 Prompts, configurable 1-4

9650-0308-01 Rev. A

ZOLL AED Plus Service Manual

Index–1

INDEX R

Radio frequency interference vi Repairs vii Returns vii RFI vi

Tests, maintenance 1-1 checklist A-1 prerequisites 1-2 Troubleshooting 2-1 common problems 2-2 V

S

Safety considerations vi Self-test, manual 1-9 Service, technical vii Simulator iv, 1-2 Software, ZOLL Administration 1-2, 2-4

Voltage warning vi W

Warranty vi Wording, configurable 1-4

T Z

Technical Service vii ZOLL Administration Software (ZAS) 1-2, 2-4

Index–2

www.zoll.com

9650-0308-01 Rev. A