TURKEY REPUBLIC MINISTRY OF ENVIRONMENT AND FORESTRY TURKISH STATE METEOROLOGICAL SERVICE DIRECTORY OF METEOROLOGICAL INSTRUMENTS AND CALIBRATION

Calibration Center - Ankara 2010

Calibration Center

At

the

Calibration

Center: Automated

Weather Observing System's sensors that is installed to all around the Türkiye are calibrated

by

applying

the

17025

standards.

Calibration Center is accredited for the calibrations

of

tempetature,

relative

humuidity, air pressure and wind speed sensors. Calibration Center - Ankara 2010

CALIBRASTION CENTER ORGANIZATIONAL CHART ADMINISTRATION AND FINANCE DEPARTMENT

Quality Manager

Documentation Unit

Director of Meteorological Instruments and Calibration Branch

Senior Chief of Calibration Laboratories

Instrument Acceptance, Registration , and Archive

Wind Speed Calibration Lab. Pressure Calibration Lab.

Temperature & Humidity Calibration Lab. Radiation Calibration Lab. Precipitation Calibration Lab.

Calibration Center - Ankara 2010

DIRECTORY OF METEOROLOGICAL INSTRUMENTS and CALIBRATION PRECIPITATION

WIND

Calibration Center operates under Meteorological Instruments and Calibration Division.

TEMP&HUM SOLAR RADIATION

Mr. Nur DOĞAN Director of Meteorological Instruments and Calibration Division

PRESSURE Calibration Center - Ankara 2010

QUALITY MANAGER On behalf of the management:

Responsible for the operation of the Calibration Center in accordance with

the

rules

of

the

17025

standards. Mr. Yusuf Ö. CANDEMİR

QUALITY MANAGEMENT PERSONNEL NAME & SURNAME

POSITION

CADRE

GRADUATION

YUSUF ÖZCAN CANDEMİR

QUALITY MANAGER

ENGINEER

ELECTRONICS

AHMET APAYDIN

QUALITY MANAGER ASSISTANT

ENGINEER

MECHANICAL

Calibration Center - Ankara 2010

CALIBRATION CENTER UNITS

-Instrument Acceptance, Registration and Archive, -Temperature & Humidity Calibration

Laboratory, -Pressure Calibration Laboratory, -Wind Speed Calibration Laboratory, Mr. Tekin SOYLU Senior chief of the Laboratories

-Radiation Calibration Laboratory, -Precipitation Calibration Laboratory.

Calibration Center - Ankara 2010

CALIBRATION CENTER PERSONNEL CHART NO

NAME & SURNAME

POSITION

CADRE

GRADUATION

1

TEKİN SOYLU

MANAGER FOR THE LABORATORIES

ENGINEER

ELECTRONICS

2

ZAFER TURGAY DAĞ

CHIEF OF RADIATION LAB.

ENGINEER

ELECTRONICS

3

AŞIR YASDUR

RADIATION LAB. PER.

TECHNICIAN

ELECTRONICS

4

Y. ÖZCAN CANDEMİR

CHIEF OF WIND SPEED LAB.

ENGINEER

ELECTRONICS

5

TUĞRUL ÇUKURLUOĞLU

WIND SPEED LAB. PER.

TECHNICIAN

6

HAMZA AYDIN CESSUR

CHIEF OF TEMP&HUMİDİTY LAB.

ENGINEER

7

YUSUF KARAÇOR

TEMPERATURE&HUMİDİTY LAB. PER.

TECHNICIAN

ELECTRONICS

8

AYDIN ÜREN

TEMPERATURE&HUMİDİTY LAB. PER.

TECHNICIAN

ELECTRONICS

9

İBRAHİM ARPA

CHIEF OF PRESSURE LAB.

ENGINEER

MECHANICAL

10

İRFAN KAYAPINAR

PRESSURE LAB. PER.

TECHNICIAN

ELECTRICAL

11

MUSA YER ALTI

PRESSURE LAB. PER.

TECHNICIAN

METAL

12

SERPİL SAYAR

CHIEF OF PRECIPITATION LAB.

ENGINEER

13

MELEK DOĞAN

PRECIPITATION LAB. PER.

TECHNICIAN

14

NURTEKİN ULUCAN

INSTRUMENT ACCEPTANCE®ISTRATION AND ARCHIVE

STAFF

15

MELTEM ŞEN

INSTRUMENT ACCEPTANCE®ISTRATION AND ARCHIVE

TECHNICIAN

Calibration Center - Ankara 2010

ELECTRICAL INDUSTRY

METEOROLOGY ELECTRICAL HIGH SCHOOL ELECTRICAL

INSTRUMENT ACCEPTANCE & REGISTRATION AND ARCHİVE

Instruments

accepted

to

the

Calibration Center are registered and

related

documents

are

archived in accordance with the rules of 17025 standards.

Calibration Center - Ankara 2010

CALIBRATION CENTER ENVIRONMENTAL CONDITIONS OF THE LABRATORIES

Environmental conditions of the Laboratories are controlled by using precision air-conditioner. The measured data of temperature and humuidity are automatically displayed and stored. LABORATORY

TEMPERATURE

HUMIDITY

PRESSURE

Pressure

23 °C ± 1 °C

45%RH ± 10%RH

Atmospheric pressure

Temperature & Humidity

23 °C ± 2 °C

45%RH ± 15%RH ---

Wind Speed

23 °C ± 10 °C

50%RH ± 30%RH ---

Radiation

23 °C ± 2 °C

45%RH ± 10%RH ---

Precipitation

23 °C ± 2 °C

45%RH ± 10%RH ---

Calibration Center - Ankara 2010

CALIBRATION CENTER ENVIRONMENTAL CONDITIONS OF THE LABRATORIES

Devices to control the laboratory environmental conditions:

Liebert Hiross HPM Precision Air conditioner

Fluke Hart Scientific 1620A Measuring, displaying and recording. Calibration Center - Ankara 2010

TEMPERATURE & HUMIDITY LABORATORY

The

thermometer

types

to

calibrated: 1. Liquid glass thermometers, 2. Resistance thermometers, 3. Analog output thermometers, 4. Thermometers with display. All types of relative humidity sensors are calibrated.

Calibration Center - Ankara 2010

be

TEMPERATURE & HUMIDITY LABORATORY

Scope of Accreditation Quantity, Instrument to be calibrated

TEMPERATURE Resistance Thermometers

Liquid Glass Thermometer Analog Output Thermometer

Measurement Range

Best Measurement Capability (k=2) (±)

Remarks

Water Triple Point 0,01 0C

0,014 °C

Using Triple Point

-40 0C ile +50 0C

0,04 0C

Using Liquid Bath

-40 0C ile +50 0C

0,13 0C

Using Climate Chamber

-40 0C ile +50 0C

0,06 0C

Using Liquid Bath

0,05 °C

Using Liquid Bath

0,13 °C

Using Cabin air-conditioning

0,05 °C

Using Liquid Bath

0,13 0C

Using Climate Chamber

%10 - %80

% 1,2

Using Humidity Generator

%20 - %80

% 2,5

Using Climate Chamber

-40 0C ile +50 0C

Thermometers with display -40 0C ile +50 0C RELATIVE HUMIDITY Relative Humidity Sensors

Calibration Center - Ankara 2010

TEMPERATURE & HUMIDITY LABORATORY

Reference devices for relative humidity calibration:

Vaisala HMP76

Thunder Scientific 2500 Humidity Generator. Operation range: -10˚C / +70˚C 10%RH / 90%RH Calibration Center - Ankara 2010

TEMPERATURE & HUMIDITY LABORATORY

Climate provider for Relative Humidity Calibration:

Weiss SB22 500 Climate Chamber. Operation Range:

-50˚C / +50˚C 10%RH / 90%RH

Calibration Center - Ankara 2010

TEMPERATURE & HUMIDITY LABORATORY

Reference devices for Temperature Calibration:

Isotech 909Q Standard Platinum

Hart Scientific 5615 Platinum

Resistance Thermometer

Resistance Thermometer

Calibration Center - Ankara 2010

TEMPERATURE & HUMIDITY LABORATORY

Climate provider for Temperature Calibration:

Weiss SB22 500 Fluke 7381 Liquid bath.

Climate Chamber. Operation Range: -50˚C / +50˚C

Operation range:

10%RH / 90%RH Calibration Center - Ankara 2010

-80˚C /+110˚C

TEMPERATURE & HUMIDITY LABORATORY

Secondary devices for Temperature & Humidity Laboratory:

Fluke PRT measuring unit, Zeno 3200 Data Logger and Fluke 8846A

6 ½ digit multimeter are used to measure the data.

Calibration Center - Ankara 2010

TEMPERATURE & HUMIDITY LABORATORY

Calibration at The Triple Point of Water (TPW) and Ice Point Calibration :

Calibration Center - Ankara 2010

PRESSURE CALIBRATION LABORATORY

Pressure

sensors

types

to

be

calibrated: 1. Frequency output barometers, 2. Digital output barometers, 3. Barometers with digital display,

4. Mechanical barometers.

Calibration Center - Ankara 2010

PRESSURE CALIBRATION LABORATORY

Scope of Accreditation Best Quantity, Measurement Instrument Measurement Measurement Capability to be Range Conditions (k=2) (±) calibrated Pressure Absolute Pressure Barometers

750 mbar – 1050mbar

Pressure Cabin (Gas)

0,12 mbar

Calibrator (Gas)

0,08 mbar

Calibration Center - Ankara 2010

Remarks

Procedures are prepared with DKD R6.1 , EURAMET/cg17/v.01 and OIML R 97

PRESSURE CALIBRATION LABORATORY

Reference device for Pressure Calibration:

Parosceintific 785 digiquartz barometer

Calibration Center - Ankara 2010

PRESSURE CALIBRATION LABORATORY

Climate providers for Pressure Calibration:

Fluke PPC4-ui A100K Calibrator. Dry air are used (Azoth gas).

Calibration Center - Ankara 2010

Pressure Chamber

PRESSURE CALIBRATION LABORATORY

Secondary devices of Pressure Laboratory:

Druck DPI 145

Measurement device

Measurement device for frequency output barometer.

for digital output

barometer. Calibration Center - Ankara 2010

PRESSURE CALIBRATION LABORATORY

Secondary devices of the Pressure Laboratory:

Paroscientific

765

Ruska 6200 barometer.

digiquartz barometer. Calibration Center - Ankara 2010

WIND SPEED CALIBRATION LABORATORY

Wind Speed sensor types to be calibrated (Cup, hotwire, ultrasonic, pitot, propeller): 1. Electronics wind speed sensors,

2. Mechanical wind speed sensors.

Calibration Center - Ankara 2010

WIND SPEED CALIBRATION LABORATORY

Scope of Accreditation

Quantity, Instrument to be calibrated

Best Measurement Measurement Measurement Range Conditions Capability (k=2) (±)

WIND SPEED SENSORS (REFERENCE IS ULTRASONICS)

2,5 m/s - 30 m/s

WIND SPEED SENSORS (REFERENCE IS PITOT TUBE MICROMANOMETER)

3 m/s - 34 m/s

Calibration Center - Ankara 2010

Remarks

±(0,11+%0,4xV) COMPARISON METHOD m/s

± (0,05 + %0,7xV) m/s

COMPARISON METHOD

WIND SPEED CALIBRATION LABORATORY

Reference devices for Wind Speed Calibration:

Furness Control FCO510 Micromanometer and Pitot tube.

Thies 2D 4.3810.00.340 Ultrasonic wind speed sensor

Calibration Center - Ankara 2010

WIND SPEED CALIBRATION LABORATORY

Medium provider for Wind Speed Calibration:

Wind Speed Tunnel is the Göttinger type, closed loop and open test area ( 0 – 40 m/s).

Calibration Center - Ankara 2010

WIND SPEED CALIBRATION LABORATORY

Devices of the Wind Speed Calibration Laboratory:

Hotwire Testo 435

Druck PTX7517-1 Pressure (600–1100mbar)

Wind Speed Sensor.

PR 5333A Temperature (0-40ºC) sensors. Calibration Center - Ankara 2010

WIND SPEED CALIBRATION LABORATORY

Devices of the Wind Speed Calibration Laboratory:

Vaisala WMS302M and WAA151, Thies Clima 4.3303.22.007, Qualimetrics

ECN4738,

Siap

handel

type,

Vector

İnst.

A100M/HE-1 and Lastem DN002 Wind Speed sensors are used. Calibration Center - Ankara 2010

RADIATION CALIBRATION LABORATORY

Pyranometers are calibrated at the Radiation Calibration Laboratory under the constant irradiance and new sensitivity coefficients for the calibrated sensors are calculated.

Calibration Center - Ankara 2010

RADIATION CALIBRATION LABORATORY

Scope of the Calibration

Quantity, Instrument to be calibrated

Measure ment Range

Measurement Conditions

Best Measurement Capability (k=2) (±)

Comparison with CPM22 (used as a reference) by 500W/m²– Pyranometer using constant 700W/m² light irradiance indoor calibration

Calibration Center - Ankara 2010

±2W/m²

Remarks

COMPARISON METHOD

RADIATION CALIBRATION LABORATORY

Reference device for Pyranometer Calibration:

Kipp&zonen CMP22 Pyranometer

Calibration Center - Ankara 2010

RADIATION CALIBRATION LABORATORY

Medium provider for Pyranometers Calibration:

Kipp&Zonen Calibration Facility for Pyranometers. Indoor Pyranometer Calibration System under constant irradiance light (500W/m² - 700W/m²). Calibration Center - Ankara 2010

RADIATION CALIBRATION LABORATORY

Devices of the Pyranometers Calibration Laboratory:

Keithley 2000 6 ½

Kipp&Zonen CMP21

digit multimeter

Pyranometer

Calibration Center - Ankara 2010

PRECIPITATION CALIBRATION LABORATORY

Precipitation (Rain Gauge) sensors types to be calibrated: 1. Tipping bucket rain gauges,

2. Weighing type rain gauges.

Calibration Center - Ankara 2010

PRECIPITATION CALIBRATION LABORATORY

Scope of calibration

Quantity, Instrument to be calibrated

Measure ment Range

Precipitation 1– 4000 gr (Rain Gauge)

Measurement Conditions

Best Measurement Capability (k=2) (±)

Precision Electronic Balances And peristaltic pump

Calibration Center - Ankara 2010

%1

Remarks

COMPARISON METHOD

PRECIPITATION CALIBRATION LABORATORY

Reference devices for Rain Gauges Calibration:

Mettler Toledo XP4002S

Baykon F1 Class Weigths

Precision Electronic Balances

1 gr / 2 kg

800 gr. / 4100 gr. Calibration Center - Ankara 2010

PRECIPITATION CALIBRATION LABORATORY

Medium provider for Rain Gauges Calibration:

Precipitation

simulation

is

prepared by using the Watson Marlow

520Du

peristaltic

pump which calculates the necessary amount of water for the desired amount of Rain İntensity.

Calibration Center - Ankara 2010

PRECIPITATION CALIBRATION LABORATORY

Secondary devices of the Precipitation Laboratory:

Zeno 3200 Data Logger Caps different

and

tubes

type

dimensions. Calibration Center - Ankara 2010

with and

THANK YOU FOR YOUR INTEREST

Calibration Center - Ankara 2010